Skip to main content

MCST Banner Image

Latest News

Terra MODIS Collection 6 Level 1B products will be reprocessed to correct for the long term drift observed in the reflectance trending of band 5. More details regarding the reprocessing schedule can be found here: http://ladsweb.nascom.nasa.gov/news/news080414.html

Also, information regarding this calibration change can be found here.
https://mcst.gsfc.nasa.gov/content/collection-6-information

The presentations from the MODIS Science Team Meeting, April-2014 are available at the following location: http://modis.gsfc.nasa.gov/sci_team/meetings/201404/

Terra experienced an anomaly for B36/D7 beginning on 2014109.1405. Unlike previous instances of this unknown electronics/formatter anomaly, the detector behavior has not returned to normal.After 2 weeks continually being in the anomaly state, the decision was made to change the QA flag for this detector from 'Noisy' to 'Inoperable'. All Band 36 detectors are currently flagged as Noisy.

This QA change will be in effect from the LUT versions 5.0.46.73 (Collection 5) and 6.1.14.26 (Collection 6)

May 4, 2014 marked the 12 year anniversary of Aqua launch. The MODIS instrument on-board the Aqua spacecraft has well surpassed its design lifetime, but continues to provide valuable datasets for studies of Earth's land, ocean and atmosphere systems.

MCST arranged a workshop (via telecon) to discuss the status, impact on science products and way forward in regards to the Aqua CFPA temperature issue. The presentation can be found in the following location: https://mcst.gsfc.nasa.gov/mcst-meetings